Measurement Principle
The turbidity sensor is based on the combined infrared absorption and scattered light method. Infrared light emitted by the light source is scattered by the turbidity in the sample, then converted into an electrical signal by the photoelectric detector. After analog and digital signal processing, the turbidity value of the sample is obtained.
Features
l The sensor uses dual-beam infrared scattered light photometer detection technology, offering good repeatability and stability;
l Application of the ISO7027 standard method (infrared light scattering technology) eliminates the influence of sample color;
l Built-in self-diagnostic function ensures data accuracy;
l Optional automatic cleaning brush function significantly reduces sensor maintenance;
l Digital sensor with strong anti-interference capability and long transmission distance;
l Standard digital signal output enables integration and networking with other devices without a transmitter;
l Easy and quick on-site installation of the sensor, supporting plug-and-play;
l Simple to use with reliable results; interface operation instructions reduce operational errors;
l Equipped with a visual alarm system providing important warning functions;
l The controller supports wall mounting/panel mounting/pipeline clamping installation;
l Reverse polarity protection for sensor power supply;
l Protection against incorrect connection of RS485 A/B terminals to power supply.
Typical Applications
l Online turbidity monitoring at water plant inlets, sedimentation tanks, and other stages;
l Online turbidity monitoring in sewage plants, various industrial process water, and wastewater treatment processes.
Sensor Technical Parameters:
Measurement Range | Turbidity: 0.01-100 NTU, 0.01-4000 NTU |
Measurement Accuracy | Less than ±2% of the measured value or ±0.1NTU, whichever is greater |
Repeatability | ±2% |
Resolution | 0.01~0.1NTU, depending on the range |
Pressure Range | ≤0.4Mpa |
Flow Velocity | ≤2.5m/s, 8.2ft/s |
Calibration | Sample calibration, slope calibration |
Main Materials of Sensor | Body: SUS316L (standard version), titanium alloy (seawater version); Upper and lower covers: POM; Cable: PUR |
Storage Temperature | -15 to 50℃ |
Operating Temperature | 0 to 45℃ (non-freezing) |
Dimensions | Diameter 60mm * Length 261mm |
Weight | 1.65KG |
Protection Rating | IP68/NEMA6P |
Cable Length | Standard 10-meter cable, extendable to 100 meters |
DS-115T Controller Technical Parameters:
Display Output | 128*64 graphic dot matrix LCD with strong LED backlight for operation under direct sunlight |
Power Supply | AC power: 85-500VAC (50/60HZ); DC power: 9~36VDC |
Main Materials | PA66+GF25+FR (upper cover); Powder-coated aluminum alloy (lower shell) |
Output | Three channels of 4-20mA |
Relay | Three relays, with programmable response parameters and values |
Digital Communication | Equipped with MODBUS RS485 communication function for real-time transmission of measured values |
Storage Temperature | -20 to 70℃ |
Operating Temperature | -15 to 60℃ |
Waterproof Rating | IP65/NEMA4X |
Dimensions | 145mm*125mm*162mm (L*W*H) |
Weight | 1.35KG |
| Industry Category | Measurement-Analysis-Instruments |
|---|---|
| Product Category | |
| Brand: | 西安东升 |
| Spec: | DS-TU800 |
| Stock: | 9999 |
| Manufacturer: | |
| Origin: | China / Shaanxi / Xianshi |